Two procedures to improve the performance of the Hach EZ test kit for quantifying inorganic arsenic concentrations in drinking water have been investigated. In the first, a digital image of the colored spot formed on the test strip, obtained with a flat-bed scanner was analyzed, by the computer program Colors, for the R, G, and B values. Calibrations were constructed by plotting the B values as a function of concentration. Agreement between the experimentally determined B-values and those of the printed chart was only obtained by either increasing the reaction time (to 40 min) or increasing the reaction temperature. The precision as a function of concentration was quantified. A comparison with previously estimated values for visual comparison of the colours, showed that the improved precision of the digital analysis would produce fewer false positive and fewer false negative results at the important threshold values of 10 and 50 μg L~(-1). By running the test for 24 h, improved performance at the low concentration (around 10 μg L~(-1)) end of the response scale was obtained.
机译:已经研究了两种改进Hach EZ测试套件性能的方法,以量化饮用水中的无机砷浓度。首先,通过计算机程序“色彩”分析在平板扫描仪上获得的形成在测试条上的色斑的数字图像的R,G和B值。通过将B值绘制为浓度的函数来构建校准。仅通过增加反应时间(至40分钟)或增加反应温度,才能获得实验确定的B值与印刷图表之间的一致性。精确度是浓度的函数。与先前估计的颜色进行视觉比较的值的比较表明,数字分析的改进精度将在10和50μgL〜(-1)的重要阈值下产生更少的假阳性和更少的假阴性结果。通过将测试运行24小时,在响应规模的低浓度(约10μgL〜(-1))端获得了改进的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号