首页> 外文期刊>Analytical methods >Analysis of four anticoagulants rodenticides in foodstuff and body fluids by high-performance liquid chromatography with photodiode-array detector
【24h】

Analysis of four anticoagulants rodenticides in foodstuff and body fluids by high-performance liquid chromatography with photodiode-array detector

机译:高效液相色谱-光电二极管阵列检测器分析食品和体液中的四种抗凝灭鼠剂

获取原文
获取原文并翻译 | 示例

摘要

A sensitive high-performance liquid chromatography (HPLC) method was developed for the determination of four anticoagulant rodenticide residues (warfarin, coumatetralyl, crimidine, and bromadiolone) in foodstuffs and body fluids. The conditions of accelerated solvent extraction (ASE) and HPLC were optimized. Good linearity was achieved with a correlation coefficient (r) of 0.9999. The limits of quantification for the four rodenticides ranged from 2.2 μg kg~(-1) (crimidine) to 13 μg kg~(-1) (bromadiolone) for solid samples and from 0.02 μg mL~(-1) (crimidine) to 0.15 μg mL~(-1) (bromadiolone) for liquid samples. The intra- and inter-day relative standard deviations (RSDs) of the four analytes at two spiked levels were in the range of 0.3-1.1% and 0.5-1.6%, respectively. The mean recoveries for warfarin, coumatetralyl, crimidine, and bromadiolone at the two spiked levels were in the range of 94-106% with RSDs of 0.3-1.1% for the rice and corn samples, and 87-108% with RSDs of 0.3-0.8% for the whole blood and urine samples. The proposed method has characteristics of high precision, sensitivity and accuracy as well as low consumption of the reagents. The proposed HPLC method was successfully applied for the analysis of the four anticoagulant rodenticides in rice, corn, whole blood, and urine samples.
机译:建立了一种灵敏的高效液相色谱(HPLC)方法,用于测定食品和体液中的四种抗凝血灭鼠剂残留物(华法林,香豆基,吗啡和溴苯丙隆)。优化了加速溶剂萃取(ASE)和HPLC的条件。相关系数(r)为0.9999,可获得良好的线性度。四种杀虫剂的定量限范围从固体样品的2.2μgkg〜(-1)(克里米定)到13μgkg〜(-1)(溴可乐酮)和0.02μgmL〜(-1)(crimidine)至液体样品为0.15μgmL〜(-1)(溴苯丙隆)在两个加标水平下,四种分析物的日内和日间相对标准偏差(RSD)分别在0.3-1.1%和0.5-1.6%的范围内。在两个加标水平下,华法林,香豆碱,嘧啶和溴苯丙隆的平均回收率在94-106%的范围内,大米和玉米样品的RSD为0.3-1.1%,在87-8%的RSD为0.3-全血和尿液样本的0.8%。该方法具有精度高,灵敏度高,准确度高,试剂消耗低的特点。所提出的高效液相色谱方法已成功地用于分析大米,玉米,全血和尿液样品中的四种抗凝灭鼠剂。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号