A novel flow injection chemiluminescence method for the determination of ractopamine in urine by using trivalent copper

机译:三价铜流动注射化学发光法测定尿样中莱克多巴胺。

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, a novel chemiluminescence method was proposed for the determination of ractopamine by combining with flow injection technology. The chemiluminescence system was based on oxidation of luminol by diperiodatocuprate(III) in the presence of ractopamine. The diperiodatocuprate(III) was obtained by complexation of copper in the trivalent oxidation state with periodate in a strong alkaline medium. Hereby under the optimum conditions, the relative CL intensity was proportional to the concentration of ractopamine over the range 1.0 x 10~(-9) g mL~(-1) to 1.0 x 10~(-6) g mL~(-1), with a limit of detection (3σ) of 3.4 x 10~(-10) g mL~(-1). The relative standard deviation (R.S.D.) was 2.1% for 5.0 x 10~(-8) g mL~(-1) ractopamine (n = 7). The sensitive method was successfully applied for the determination of ractopamine (ng mL~(-1)) in urine samples. The possible mechanism of the CL reactions was also discussed.
机译:结合流动注射技术,提出了一种新的化学发光法测定莱克多巴胺。化学发光系统是基于在莱克多巴胺存在下二硬脂酸铜(III)对鲁米诺的氧化作用。通过在强碱性介质中将三价氧化态的铜与高碘酸盐络合获得二酮十二碳二氢硼酸盐(III)。因此,在最佳条件下,相对CL强度与莱克多巴胺的浓度成正比,范围为1.0 x 10〜(-9)g mL〜(-1)至1.0 x 10〜(-6)g mL〜(-1) ),检出限(3σ)为3.4 x 10〜(-10)g mL〜(-1)。对于5.0 x 10〜(-8)g mL〜(-1)莱克多巴胺(n = 7)的相对标准偏差(R.S.D.)为2.1%。灵敏的方法已成功应用于尿样中莱克多巴胺(ng mL〜(-1))的测定。还讨论了CL反应的可能机理。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号