A flow injection system, comprising a biosensor based on a screen-printed carbon electrode containing Meldola's Blue-Reinecke salt coated with glucose dehydrogenase, for the measurement of glucose

机译:流动注射系统,包括基于丝网印刷碳电极的生物传感器,该碳电极包含涂覆有葡萄糖脱氢酶的梅尔多拉的蓝-雷内克盐,用于测量葡萄糖

获取原文
获取原文并翻译 | 示例
           

摘要

A biosensor for the measurement of glucose in serum has been developed, based on a screen-printed carbon electrode modified with Meldola's Blue-Reinecke salt, coated with the enzyme glucose dehydrogenase (from Bacillus sp.), and nicotinamide adenine dinucleotide coenzyme (NAD~+). A cellulose acetate layer was deposited on top of the device to act as a permselective membrane. The biosensor was incorporated into a commercially available, thin-layer, amperometric flow cell operated at a potential of only +0.05 V versus Ag/AgCl. The mobile phase consisted of 0.2 M phosphate buffer (pH 7.0) containing 0.1 M potassium chloride solution, and a flow rate of 0.8 ml min~(-1) was used throughout the investigation. The biosensor response was linear over the range of 0.075-30 mM glucose, with the former representing the detection limit. The precision of the system was determined by carrying out 20 repeat injections of a 5-mM glucose standard, and the calculated coefficient of variation was 3.9%. It was demonstrated that this biosensor system could be applied to the direct measurement of glucose in serum without pretreatment. Therefore, this would allow high-throughput analysis, at low cost, for this clinically important analyte.
机译:基于用Meldola的Blue-Reinecke盐修饰的丝网印刷碳电极,并涂覆了葡萄糖脱氢酶(来自Bacillus sp。)和烟酰胺腺嘌呤二核苷酸辅酶(NAD〜 +)。乙酸纤维素层沉积在该装置的顶部上以充当渗透选择膜。将生物传感器并入市售的薄层安培流通池中,该池相对于Ag / AgCl的电势仅为+0.05V。流动相由含有0.1 M氯化钾溶液的0.2 M磷酸盐缓冲液(pH 7.0)组成,整个研究过程中使用的流速为0.8 ml min〜(-1)。在0.075-30 mM葡萄糖范围内,生物传感器响应呈线性,前者表示检测极限。通过重复注射20次5-mM葡萄糖标准品来确定系统的精度,计算出的变异系数为3.9%。已证明该生物传感器系统无需预处理即可用于直接测量血清中的葡萄糖。因此,这将允许以低成本对这种临床上重要的分析物进行高通量分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号