首页> 外文期刊>Analytical and bioanalytical chemistry >Application of an in situ plated lead film electrode to the analysis of testosterone by adsorptive stripping voltammetry
【24h】

Application of an in situ plated lead film electrode to the analysis of testosterone by adsorptive stripping voltammetry

机译:原位镀铅膜电极在吸附溶出伏安法分析睾丸激素中的应用

获取原文
获取原文并翻译 | 示例

摘要

Adsorptive stripping voltammetry is a very sensitive and selective method for determination of drugs and organic substances in biological fluids. We have shown that determination of testosterone by adsorptive stripping voltammetry is possible using a lead film electrode. The lead film plating process and accumulation of testosterone were performed simultaneously from an acetate buffer solution of pH=5.2 at a potential of -1.1 V. The measurements were carried out in undeaerated solutions. The detection limit was 9x10(-9) mol L-1 for an accumulation time of 120 s; the relative standard deviation for 1x10(-7) mol L-1 testosterone was 3.8%. The proposed voltammetric procedure for determination of testosterone could be applied to its determination in a pharmaceutical preparation and human urine samples directly without any separation steps.
机译:吸附溶出伏安法是一种非常灵敏和选择性的测定生物液体中药物和有机物的方法。我们已经表明,使用铅膜电极通过吸附溶出伏安法测定睾丸激素是可能的。在pH值为5.2的醋酸盐缓冲溶液中以-1.1 V的电位同时进行铅膜镀覆过程和睾丸激素的积累。测量是在未脱气的溶液中进行的。检测限为9x10(-9)mol L-1,累积时间为120 s; 1x10(-7)mol L-1睾丸激素的相对标准偏差为3.8%。提议的用于测定睾丸激素的伏安法可以直接用于药物制剂和人尿液样品中的测定,而无需任何分离步骤。