摘要

The construction and general performance of new valproate-selective electrodes based on manganese(III) tetraphenylporphyrin [Mn(III)TPP-Cl], as an ionophore, are presented. The ionophore was incorporated into PVC and ceramic membranes (sol-gel) based on methyltriethoxysilane. The influence of membrane composition and pH and the effect of lipophilic cationic and anionic additives in PVC membranes were investigated concerning their influence on the slope, response time, selectivity and lifetime of the electrodes. The PVC membrane without additive and the sol-gel membrane presented slopes and practical limits of detection of -60.8 mV dec(-1) and 5x10(-6) mol l(-1) and -60.3 mV dec(-1) and 1x10(-4) mol l(-1), respectively. The sol-gel membranes displayed higher selectivity for valproate when compared with PVC membranes. These two types of electrodes were coupled to a sequential-injection analysis (SIA) system for the direct determination of valproate in pharmaceutical formulations. The association of Mn(III)TPP-Cl with the sol-gel support inserted in a SIA system provided potentiometric sensors with an analytical range of 1x10(-3)-5x10(-2) mol l(-1), with a sample rate of 55 samples per hour and a sample and carrier consumption of 140 and 2,500 mu l per determination, respectively.
机译:介绍了以锰(III)四苯基卟啉[Mn(III)TPP-Cl]为离子载体的新型丙戊酸选择电极的结构和一般性能。将离子载体掺入PVC和基于甲基三乙氧基硅烷的陶瓷膜(溶胶-凝胶)中。研究了膜组成和pH的影响以及亲脂性阳离子和阴离子添加剂对PVC膜的影响,研究了它们对电极的斜率,响应时间,选择性和寿命的影响。没有添加剂的PVC膜和溶胶-凝胶膜呈现的斜率和实际检测极限为-60.8 mV dec(-1)和5x10(-6)mol l(-1)和-60.3 mV dec(-1)和1x10 (-4)摩尔(-1)。与PVC膜相比,溶胶-凝胶膜对丙戊酸具有更高的选择性。将这两种类型的电极连接到顺序注射分析(SIA)系统上,以直接测定药物制剂中的丙戊酸盐。 Mn(III)TPP-Cl与插入SIA系统中的溶胶-凝胶支持物的结合为电位传感器提供了分析范围为1x10(-3)-5x10(-2)mol l(-1)的样品每小时测定55个样品,每次测定的样品和载体消耗分别为140和2500μl。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号