首页> 外文期刊>Biomedical Chromatography: An International Journal Devoted to Research in Chromatographic Methodologies and Their Applications in the Biosciences >A novel approach for concurrent quantitation of glutathione, glutathione disulfide, and 2-hydroxyethylated glutathione in lungs of mice exposed to ethylene oxide, using liquid chromatography-positive electrospray tandem mass spectrometry
【24h】

A novel approach for concurrent quantitation of glutathione, glutathione disulfide, and 2-hydroxyethylated glutathione in lungs of mice exposed to ethylene oxide, using liquid chromatography-positive electrospray tandem mass spectrometry

机译:使用液相色谱-阳性电喷雾串联质谱法同时测定暴露于环氧乙烷的小鼠肺中谷胱甘肽,谷胱甘肽二硫和2-羟乙基化谷胱甘肽的新方法

获取原文
获取原文并翻译 | 示例
           

摘要

Glutathione (GSH), glutathione disulfide (GSSG) and 2-hydroxyethylated glutathione (HESG) are important biomarkers for exploring the genotoxicity mechanism of ethylene oxide (EO) or ethylene in vivo. A liquid chromatography-tandem mass spectrometry method was developed for simultaneous determination of GSH, GSSG and HESG in mouse lung tissues after inhalation exposure to EO. The lower limit of quantitation for all these biomarkers was 0.002 mu g/mL. The linearity of the calibration curves for all analytes was >0.998. The intra-day assay precision relative standard deviation (RSD) values for quality control samples for all analytes were 12.8% with accuracy values ranging from 87.2 to 113%. The inter-day assay precision (RSD) values for all analytes were 13.1% with accuracy values ranging from 86.9 to 103%. This method was applied to concurrently determine the levels of GSH, GSSG and HESG in lung samples isolated from mouse after 4-week inhalation exposure to EO at 0, 10, 50, 100 and 200 ppm. Copyright (c) 2015 John Wiley & Sons, Ltd.
机译:谷胱甘肽(GSH),二硫谷胱甘肽(GSSG)和2-羟乙基化谷胱甘肽(HESG)是探索体内环氧乙烷(EO)或乙烯的遗传毒性机理的重要生物标记。建立了一种液相色谱-串联质谱法,用于同时测定暴露于EO后的小鼠肺组织中的GSH,GSSG和HESG。所有这些生物标志物的定量下限为0.002μg / mL。所有分析物的校准曲线的线性> 0.998。所有分析物的质控样品的日内测定精度相对标准偏差(RSD)值为12.8%,准确度范围为87.2至113%。所有分析物的日间测定精度(RSD)值为13.1%,准确度范围为86.9至103%。该方法用于同时测定从小鼠吸入的0、10、50、100和200 ppm的EO暴露4周后从肺中分离得到的肺样品中GSH,GSSG和HESG的水平。版权所有(c)2015 John Wiley&Sons,Ltd.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号