Hydrogen peroxide and peracetic acid determination in waste water using a reversible reagentless biosensor

机译:使用可逆无试剂生物传感器测定废水中的过氧化氢和过氧乙酸

获取原文
获取原文并翻译 | 示例
           

摘要

During the reversible reaction between peroxidase (HRP) and peroxides,several peroxidase intermediate species,showing different molecular absorption spectra,are formed which can be used for their determination.On this basis,a reversible reagentless optical biosensor based on HRP for hydrogen peroxide and peracetic acid determinations has been developed.The biosensor (which can be used for at least 3 months and/or more than 200 measurements) is prepared by HRP entrapment in a polyacrylamide gel matrix.A mathematical model (in which optical,kinetic and transport aspects are considered) relating the measured absorbance with the analyte concentration is also presented.Both peroxides show similar responses in the sensor film.Under the recommended working conditions,the biosensor shows linear response ranges from 6 x 10~(-7) to 1.0 x 10~(-4) M using FLA mode,and from 2 x 10~(-7) to 1.5 x 10~(-5) M using continuous mode for both peroxides;the precision,expressed as R.S.D.,is about 4%.This biosensor has been applied for peroxide determination in waste water samples previously treated with peroxides.
机译:在过氧化物酶与过氧化物的可逆反应过程中,形成了几种具有不同分子吸收光谱的过氧化物酶中间体,可用于测定。在此基础上,基于HRP的可逆无试剂光学生物传感器用于过氧化氢和过氧乙酸。已开发出酸测定法。通过HRP截留在聚丙烯酰胺凝胶基质中制备生物传感器(可使用至少3个月和/或进行200多次测量)。数学模型(其中光学,动力学和运输方面考虑过的)也提出了测量的吸光度与分析物浓度之间的关系。两种过氧化物在传感器膜中都表现出相似的响应。在推荐的工作条件下,生物传感器的线性响应范围为6 x 10〜(-7)至1.0 x 10〜 (-4)M使用FLA模式,对于两种过氧化物,使用连续模式从2 x 10〜(-7)到1.5 x 10〜(-5)M;以RSD表示的精度约为4%。传感器已用于预先用过氧化物处理过的废水样品中的过氧化物测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号