A direct competitive chemiluminescent immunosensor system that exploits the competition between chloramphenicol (CAP) as an analyte and CAP-horseradish peroxidase conjugate as a tracer for binding to an anti-CAP antibody on a solid support was devised by installing a flow-through cell which was connected to an injector and a peristaltic pump inside a dark box, followed by positioning a photomultiplier tube as light detector in front of it. The anti-CAP antibody was immobilized onto positively charged Biodyne B membrane pieces by a dipping procedure. The operating conditions for the immunosensor were selected with respect to substrate composition (0.25, 13.3 and 0.66 mM for luminol, H2O2 and p-iodophenol, respectively), injection volume of the substrate solution (200 mu L) and the concentrations of antibody for immobilization (0.10 mg mL(-1)) and tracer (0.030 mg mL(-1)). At these conditions, sensor response according to analyte concentration was well fitted to a linear equation when plotted in semi-logarithmic scale, with the limit of detection for CAP of 10(-8) M. By using the immunosensor, CAP measurement in the model samples prepared from five food materials was conducted. (c) 2006 Elsevier B.V. All rights reserved.
机译:通过安装流通池设计了一种直接竞争的化学发光免疫传感器系统,该系统利用了氯霉素(CAP)作为分析物与CAP-辣根过氧化物酶偶联物作为示踪剂之间的竞争来结合抗CAP抗体在固体支持物上的竞争。连接到暗箱内的注射器和蠕动泵,然后将光电倍增管作为光检测器放置在其前面。通过浸渍方法将抗CAP抗体固定在带正电的Biodyne B膜片上。根据底物组成(鲁米诺,H2O2和对碘苯酚分别为0.25、13.3和0.66 mM),底物溶液的注入量(200μL)和固定化抗体的浓度选择免疫传感器的操作条件。 (0.10 mg mL(-1))和示踪剂(0.030 mg mL(-1))。在这些条件下,当按半对数标度绘制时,根据分析物浓度的传感器响应就很好地拟合了线性方程,CAP的检测极限为10(-8)M。通过使用免疫传感器,模型中的CAP测量进行了从五种食品原料中提取的样品。 (c)2006 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号