首页> 外文期刊>Analytical and bioanalytical chemistry >Determination of impurities in magnesium niobate by slurry introduction axially viewed inductively coupled plasma optical emission spectrometry
【24h】

Determination of impurities in magnesium niobate by slurry introduction axially viewed inductively coupled plasma optical emission spectrometry

机译:浆料引入轴向观察电感耦合等离子体发射光谱法测定铌酸镁中的杂质。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A simple preparation scheme is described for the quantitative analysis of a magnesium niobate sample using slurry introduction axially viewed inductively coupled plasma optical emission spectrometry. Relationships between the stability of slurries and the conditions, such as particle size, pH, dispersant and amount of dispersant, were investigated experimentally. The MgNb2O6 slurry sample was prepared by adding the dispersant sodium polyacrylate and agitation in an ultrasonic bath to ensure good dispersion. Under optimization of pH and amount of dispersant, an analysis of minor and trace impurities (Ba, Ca, Cr, Cu, Fe, Mn, Ni, Pb) in magnesium niobate was accomplished. Applying a paired t test, we showed that the results were in agreement at a 95% confidence level with the reference values obtained by a fusion method for a magnesium niobate sample, which verified that the calibration curves could be established by aqueous standards. Analytical results demonstrate that the factors that affected the accuracy of determination for MgNb2O6 are mainly the particle size of the sample and the stability of slurry.
机译:描述了一种简单的制备方案,用于使用浆液引入轴向观察的电感耦合等离子体发射光谱法对铌酸镁样品进行定量分析。实验研究了浆料的稳定性与条件的关系,如粒径,pH,分散剂和分散剂的量。通过添加聚丙烯酸钠分散剂并在超声浴中搅拌以确保良好的分散性来制备MgNb2O6浆料样品。在优化pH和分散剂含量的条件下,完成了铌酸镁中微量和痕量杂质(Ba,Ca,Cr,Cu,Fe,Mn,Ni,Pb)的分析。应用配对t检验,我们发现结果与铌酸镁样品的融合方法获得的参考值在95%的置信度水平上吻合,这证明校准曲线可以通过水性标准液建立。分析结果表明,影响MgNb2O6测定准确性的因素主要是样品的粒径和浆液的稳定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号