首页> 外文期刊>Analytica chimica acta >Evaluation of alternative measurement and data-processing options for enzyme-based biosensors
【24h】

Evaluation of alternative measurement and data-processing options for enzyme-based biosensors

机译:评估基于酶的生物传感器的替代测量和数据处理选项

获取原文
获取原文并翻译 | 示例
       

摘要

This paper describes an alternative measurement/data-processing approach for enzyme-based biosensors. The proposed approach, identified herein as a pseudo-equilibrium option, is designed to obtain the response corresponding to reaction of all the substrate in a small fixed volume of solution. This general approach has two important advantages relative to more conventional steady-state option. It yields constant high sensitivity over wide concentration ranges and it improves the ruggedness to changes in experimental conditions by factors of 10-fold or more relative to the steady-state option. Results obtained for an amperometric-based reactor/sensor system for glucose demonstrate constant sensitivity for glucose concentrations up to five-fold the Michaelis constant and order-of-magnitude improvements in ruggedness to changes in selected experimental variables. Regarding ruggedness, results show that the pseudo-equilibrium option is 11-, 150- and 32-fold less dependent on pH, temperature and sodium chloride concentration, respectively, than steady-state results obtained from the same data sets. Principal limitations of the implementation of the pseudo-equilibrium option reported here are the possible inconvenience of a flow system, long measurement times and more complex data-processing requirements than the steady-state option. (C) 1999 Elsevier Science B.V. All rights reserved. [References: 9]
机译:本文介绍了基于酶的生物传感器的另一种测量/数据处理方法。在本文中被标识为伪平衡选项的所提出的方法被设计为获得与在较小固定体积的溶液中所有底物的反应相对应的响应。相对于更常规的稳态选择,此通用方法具有两个重要优点。它在较宽的浓度范围内可产生恒定的高灵敏度,并且相对于稳态选择,其对实验条件变化的耐受性提高了10倍或更多倍。基于电流分析法的葡萄糖反应器/传感器系统获得的结果表明,其对葡萄糖浓度的恒定敏感性高达米氏常数的五倍,并且对所选实验变量的变化的坚固性得到了数量级的提高。关于坚固性,结果表明,与从相同数据集获得的稳态结果相比,拟平衡选择对pH,温度和氯化钠浓度的依赖性分别低11倍,150倍和32倍。此处报告的伪平衡选项的实现的主要限制是流量系统可能带来的不便,比稳态选项更长的测量时间和更复杂的数据处理要求。 (C)1999 Elsevier Science B.V.保留所有权利。 [参考:9]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号