首页> 外文期刊>Analytica chimica acta >Development of a tubular fluoride potentiometric detector for flow analysis Evaluation and analytical applications
【24h】

Development of a tubular fluoride potentiometric detector for flow analysis Evaluation and analytical applications

机译:用于流量分析的管状氟化物电位检测器的开发评估和分析应用

获取原文
获取原文并翻译 | 示例
       

摘要

In this work a construction procedure for tubular fluoride electrode to be used in flow systems is outlined.The electrode was constructed from a commercially available,LaF3 single crystal.Principal advantages of the flow detector presented include simplicity of construction,robustness,durability,low cost and easy coupling into any point of a flow manifold.Evaluation of the intrinsic working characteristics of the potentiometric detector in a low dispersion manifold is presented with respect to analytical and dynamic parameters.The constructed detector has similar working characteristics to those of the conventional fluoride electrodes,namely the detection limit,lower limit of linear response and operational pH range.The analytical usefulness of the constructed device was assessed in a flow system developed for fluoride determination in toothpaste,tablet,collutory and water samples for which the reference procedures suggest the determination of fluoride ion with a conventional ion selective electrode.
机译:在这项工作中,概述了在流动系统中使用的管状氟化物电极的构建程序。该电极由市售的LaF3单晶构建而成。该流动检测器的主要优点包括结构简单,坚固耐用,成本低廉并针对低分散歧管中的电位检测器的分析和动态参数进行了评估,该检测器具有与常规氟化物电极相似的工作特性,即检测限,线性响应的下限和可操作的pH范围。在开发的用于牙膏,片剂,普通样品和水样品中氟化物测定的流动系统中评估了所构建装置的分析实用性,其中参考程序建议用于测定常规离子选择器分离氟离子电极。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com