Conventional flow-injection manifolds with sample injection for the determination of reactive phosphorus in estuarine waters are limited by the Schlieren or refractive index (RI) effect which can cause major errors in quantification. A simple flow-injection (FI) manifold which obviates this RI error in reactive phosphorus measurement is reported. It involves the injection of acidic molybdophosphate reagent into a carrier stream of sodium chloride solution of similar refractive index, which is then sequentially merged with a sample (the salinity of which may vary widely from sample to sample) and a reductant. Despite the occurrence of sizeable blank signals, reactive phosphorus has been successfully measured in samples with salinities ranging from 0 to 34 parts per thousand using calibration standards prepared in deionised water, with a detection limit of 6 mu g l(-1) P. (C) 1997 Elsevier Science B.V.
机译:用于测定河口水中反应性磷的常规进样流路与样品进样受到Schlieren或折射率(RI)效应的限制,该效应可能导致定量分析中的重大误差。据报道,一种简单的流动注射(FI)歧管可消除活性磷测量中的RI误差。它涉及将酸性钼酸磷试剂注入具有相似折射率的氯化钠溶液的载流中,然后将其与样品(其盐度可能因样品而异)和还原剂依次合并。尽管出现了大量空白信号,但使用去离子水中制备的校准标准品已成功测定了盐度范围为0至34千分之一的样品中的活性磷,检测限为6μggl(-1)P.(C )1997爱思唯尔科学有限公司