A needle-type biosensor system was developed for rapid and simple determination of glucose levels in fish blood. The sensor comprises a needle-type hollow container (18-gauge needle), immobilized enzyme membrane and optic fiber probe with ruthenium complex. The enzyme membrane was prepared from glucose oxidase, azide-unit pendant water-soluble photopolymer and an ultra-thin dialysis membrane. The optic fiber probe was inserted into the rolled enzyme membrane placed in the needle-type hollow container. The calibration curve was linear for glucose levels in fish plasma. One assay was completed within 3 min. A good reproducibility was observed for 60 times without exchange of the enzyme membrane. The sensor was inserted into the caudal vein of fish to measure blood glucose levels. The sensor responded immediately after insertion and glucose levels could be monitored. Good correlations were observed between values determined using the sensor and conventional methods in the range of 48-157 mg dl(-1) (correlation coefficient, 0.9474). (c) 2006 Elsevier B.V. All rights reserved.
机译:开发了一种针型生物传感器系统,用于快速,简单地确定鱼血中的葡萄糖水平。该传感器包括一个针型空心容器(18号针),固定的酶膜和带有钌配合物的光纤探头。该酶膜由葡萄糖氧化酶,叠氮化物单元侧基水溶性光敏聚合物和超薄透析膜制成。将光纤探针插入放置在针型中空容器中的滚动酶膜中。鱼血浆中葡萄糖水平的校准曲线是线性的。 3分钟内完成一项测定。在不更换酶膜的情况下,观察到60次良好的重现性。将传感器插入鱼的尾静脉以测量血糖水平。插入后传感器立即响应,可以监测葡萄糖水平。在使用传感器和常规方法测定的值之间,在48-157 mg dl(-1)范围内观察到良好的相关性(相关系数为0.9474)。 (c)2006 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号