SEQUENTIAL INJECTION DETERMINATION OF D-GLUCOSE BY CHEMILUMINESCENCE USING AN OPEN TUBULAR IMMOBILISED ENZYME REACTOR

机译:开管固定化酶反应器化学发光顺序注射测定D-葡萄糖

获取原文
获取原文并翻译 | 示例
       

摘要

A sequential injection analysis system is described that incorporates a nylon tubular reactor containing immobilised glucose oxidase, allowing determination of D-glucose by means of subsequent luminol chemiluminescence detection of the hydrogen peroxide generated in the enzymatic reaction. The operating parameters were optimised by fractional factorial screening and response surface modelling. The linear range of D-glucose determination was 30-600 mu M, With a detection limit of 15 mu M using a photodiode detector. The sampling frequency was 54 h(-1). Lower LOD (0.5 mu M D-glucose) could be reached by using a PMT as the detector. Fermentation broth samples were determined and good recoveries were obtained. [References: 19]
机译:描述了一种顺序注射分析系统,该系统结合了包含固定葡萄糖氧化酶的尼龙管式反应器,从而可以通过随后的鲁米诺化学发光检测酶促反应中产生的过氧化氢来确定D-葡萄糖。通过分数阶乘筛选和响应面建模优化了运行参数。 D-葡萄糖测定的线性范围为30-600μM,使用光电二极管检测器的检测极限为15μM。采样频率为54 h(-1)。通过使用PMT作为检测器,可以达到较低的LOD(0.5μM D-葡萄糖)。测定发酵液样品并获得良好的回收率。 [参考:19]