A new and simple flow injection system procedure has been developed for the indirect determination of cyanide. The method is based on insertion of aqueous cyanide solutions into an on-line cadmium carbonate packed column (25%, m/m suspended on silica eel beads) and a sodium hydroxide with pH 10 is used as the carrier stream. The eluent containing the analyte as cadmiumcyanide complexes. produced from reaction between cadmium carbonate and cyanide, measured by flame atomic absorption spectrometry. The absorbance is proportional to the concentration of cyanide in the sample. The linear range of the system is up to 15 mg L-1 with a detection limit 0.2 mg L-1 and sampling rate 72 h(-1). The method is suitable for determination of cyanide in industrial waste waters with a relative standard deviation better than 1.22%. (C) 2004 Elsevier B.V. All rights reserved.
机译:已经开发了一种新的简单的流动注射系统程序,用于间接测定氰化物。该方法基于将氰化物水溶液插入在线碳酸镉填充柱(25%,m / m悬浮在硅胶鳗珠上)中,并将pH 10的氢氧化钠用作载流。含有分析物的洗脱液为氰化镉络合物。由碳酸镉和氰化物之间的反应产生的产物,通过火焰原子吸收光谱法测量。吸光度与样品中氰化物的浓度成正比。系统的线性范围最大为15 mg L-1,检测极限为0.2 mg L-1,采样率为72 h(-1)。该方法适用于工业废水中氰化物的测定,相对标准偏差优于1.22%。 (C)2004 Elsevier B.V.保留所有权利。