摘要

New 20 and 30 mm light path lenght twin flow-cells were designed for the simultaneous flow-injection spectrophotometric determination of trace amounts of copper and iron in patients sera.Hence,utilization of the twin flow-cell enabled the sensitive determinations.Copper and iron can be determined simultaneously and rapidly at 580 and 710 nm using water-soluble reagent 5-Br-PSAA.The limits of detections (S/N=3)with the 20 mm path length flow-cell were 0.39 mug l~(-1)for copper and 0.20mug l~(-1)for iron.The relative stadnard deviations (n=10)were .81% for 20mug l~(-1)copper and 0.52% for 20 mug l~(-1)iron.Copper and iron in five kinds of control sera were determined by the proposed method and ICP-AES.The method was applied to the determination of copper and iron in patients sera and disease diagnosis.As the sensitivity of the proposed FI system was excellent,the amount of serum sample consumed in the pretreatment was small.
机译:设计了新的20 mm和30 mm光程长双流动池,用于同时流动注射分光光度法测定患者血清中的痕量铜和铁。因此,双流动池的使用使得灵敏的测定成为可能。可以使用水溶性试剂5-Br-PSAA在580和710 nm处同时快速测定。20 mm路径长度流通池的检测限(S / N = 3)为0.39 mug l〜(-1铜)和0.20杯l〜(-1)铁的相对标准偏差(n = 10)对于20杯l〜(-1)铜为0.81%,对于20杯l〜(-1)的铁为0.52%采用该方法和ICP-AES测定了5种对照血清中的铜和铁,该方法用于患者血清中铜和铁的测定以及疾病的诊断。由于该FI系统的灵敏度极佳,预处理中消耗的血清样品量很小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号