首页> 外文期刊>Analytica chimica acta >Flow electrochemical determination of ascorbic acid in real samples using a glassy carbon electrode modified with a cellulose acetate film bearing 2,6-dichlorophenolindophenol
【24h】

Flow electrochemical determination of ascorbic acid in real samples using a glassy carbon electrode modified with a cellulose acetate film bearing 2,6-dichlorophenolindophenol

机译:使用载有2,6-二氯苯酚吲哚酚的醋酸纤维素膜修饰的玻碳电极,通过流动电化学方法测定实际样品中的抗坏血酸

获取原文
获取原文并翻译 | 示例
           

摘要

An ascorbate sensor based on a glassy carbon electrode modified with a cellulose acetate polymeric film bearing 2,6-dichlorophenolindophenol (CA/DCPI-CME) was constructed. The overall reaction obeys a catalytic regeneration mechanism (EC mechanism) and the electrochemical rate constant kf for the electrocatalytic oxidation of ascorbic acid was evaluated. The modified electrodes were mounted in a flow injection (FI) manifold, poised at +100 mV versus Ag/AgCl/3 M KCl at pH 6.5 and utilized for the determination of ascorbic acid in beverages and pharmaceuticals. Good correlation with a reference method was attained. Interferents of various molecular sizes were tested. Calibration graphs were linear over the range 0.02-1 and 0.1-6 mM ascorbic acid for CA/DCPI sensors hydrolyzed in KOH and ZnCl2 solution, respectively. The throughput was 25 samples per hour and the CV was for a 0.4 mM ascorbic acid solution 0.75 (n=14) and 1.2% (n=10) for CA/DCPI sensors hydrolyzed in KOH and ZnCl2 solution, respectively. The recovery was 92-110%. The sensors showed very good repeatability and operational stability. (C) 2000 Elsevier Science B.V. All rights reserved. [References: 28]
机译:构造了一种抗坏血酸传感器,该传感器基于玻璃碳电极,该电极由带有2,6-二氯苯酚吲哚酚(CA / DCPI-CME)的醋酸纤维素聚合物膜改性而成。整个反应遵循催化再生机理(EC机理),并评估了抗坏血酸的电催化氧化的电化学速率常数kf。将修饰的电极安装在流动注射(FI)歧管中,相对于pH 6.5的Ag / AgCl / 3 M KCl保持在+100 mV的电位,并用于测定饮料和药品中的抗坏血酸。与参考方法具有良好的相关性。测试了各种分子大小的干扰素。对于在KOH和ZnCl2溶液中水解的CA / DCPI传感器,校准曲线分别在0.02-1和0.1-6 mM抗坏血酸范围内呈线性。对于每小时在KOH和ZnCl2溶液中水解的CA / DCPI传感器,0.4毫升抗坏血酸溶液的通量为每小时25个样品,CV分别为0.75(n = 14)和1.2%(n = 10)。回收率为92-110%。传感器显示出非常好的重复性和操作稳定性。 (C)2000 Elsevier Science B.V.保留所有权利。 [参考:28]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号