Direct determination of phosphate in soil extracts by potentiometric flow injection using a cobalt wire electrode

机译:使用钴线电极通过电位流动注射直接测定土壤提取物中的磷酸盐

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A system is described for the direct potentiometric flow injection analysis of phosphates using a cobalt wire electrode. It is proposed that the electrode response to phosphate is a result of CO3(PO)(2) precipitate formation at the electrode surface. The electrode response to phosphate was found to be depend on the phthalate concentration, carrier pH, flow rate and injected volume. A linear response with a slope of -41+/-0.5 mV/decade was obtained in the concentration range of 1.0x10(-5) -5.0x10(-3) M with a detection limit of 1x10(-6) M, using a carrier solution containing 25 mM potassium hydrogen phthalate at pH 4.0. The proposed method was used for the determination of soil extracts with a fast analysis, low cost, high sensitivity and selectivity. (C) 1998 Elsevier Science B.V. [References: 28]
机译:描述了一种用于使用钴线电极对磷酸盐进行直接电位流注入分析的系统。建议电极对磷酸盐的响应是在电极表面形成CO3(PO)(2)沉淀的结果。发现电极对磷酸盐的响应取决于邻苯二甲酸酯浓度,载体pH,流速和注入量。使用1.0x10(-5)-5.0x10(-3)M的浓度范围获得线性斜率为-41 +/- 0.5 mV /十倍的线性响应,检测限为1x10(-6)M,使用载体溶液,其中溶液的pH值为25 mM邻苯二甲酸氢钾。该方法用于分析土壤中的提取物,具有分析速度快,成本低,灵敏度高和选择性高的特点。 (C)1998 Elsevier Science B.V. [参考:28]