首页> 外文期刊>Analytica chimica acta >Study of parameters associated with the determination of cadmium in lake sediments by slurry sampling electrothermal atomic absorption spectrometry
【24h】

Study of parameters associated with the determination of cadmium in lake sediments by slurry sampling electrothermal atomic absorption spectrometry

机译:泥浆采样电热原子吸收光谱法测定湖泊沉积物中镉的相关参数研究

获取原文
获取原文并翻译 | 示例
           

摘要

Cadmium concentration in lake sediments is determined by suspending the solid samples in a solution containing 5% (v/v) concentrated nitric acid and 0.1% (v/v) Triton X-100.Three modifiers were tested for the direct determination.The furnace temperature programmes and appropriate amount for each modifier were optimised to get the highest signal and the best separation between the atomic and background signals.The drying stage is performed by programming a 400 deg C temperature,a ramp time of 25s and hold time of 10s on the power supply of the atomiser.No ashing step is used and platform atomisation is carried out at 2200 C.W-Rh permanent modifier combined with conventional modifier by delivering 10 mul of 0.50% (w/v) NH_4H_2PO_4 solution was the best chemical modifier for cadmium determination.This modifier also acts as a liquid medium for the slurry,thus simplifying the procedure.Calibration is performed using aqueous standards in the 1-5 mug l~(-1) range.The optimised method gave a limit of detection of 0.56 ng ml~(-1),characteristic mass of 10.1 +- O.Spg for aqueous standard,9.6 +-0.7 pg for slurry samples containing different Cd concentrations and good precision (7.6-5.2%).The method was validated by analysing four certified reference lake sediment materials: LKSD-1,LKSD-2,LKSD-3 and LKSD-4; satisfactory recoveries were obtained (90.0-96.3%) and no statistical differences were observed between the experimental and the certified cadmium concentration.The developed methodology was used to determine cadmium in three 'real' sediment samples from lakes in the area of Wielkopolski National Park,Poland.
机译:通过将固体样品悬浮在含有5%(v / v)浓硝酸和0.1%(v / v)Triton X-100的溶液中来测定湖泊沉积物中的镉浓度,并测试了三种改性剂以直接测定。优化了温度程序和每种修饰剂的适量,以获得最高的信号以及原子和背景信号之间的最佳分离。通过设定400℃的温度,25s的斜坡时间和10s的保持时间来执行干燥阶段无需灰化步骤,平台雾化在2200 CW-Rh永久性改性剂与常规改性剂的结合下进行,通过输送10 mul 0.50%(w / v)NH_4H_2PO_4溶液是镉的最佳化学改性剂该修饰剂还可以用作浆液的液体介质,从而简化了操作过程。使用1-5杯l〜(-1)范围内的水标准液进行校准。该方法的检测灵敏度为0.56 ng ml〜(-1),特征质量为水标准品为10.1 + -O.Spg,含有不同Cd浓度且精密度为(7.6-5.2%)的浆料样品为9.6 + -0.7 pg。通过分析四种经认证的参比湖沉积物来验证:LKSD-1,LKSD-2,LKSD-3和LKSD-4;获得了令人满意的回收率(90.0-96.3%),并且实验镉浓度与标准镉浓度之间没有统计学差异。开发的方法用于测定Wielkopolski国家公园附近三个湖泊的``真实''沉积物样品中的镉。波兰。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号