DIRECT DETERMINATION OF THE TRANSFER COEFFICIENT FROM CYCLIC VOLTAMMETRY - ISOPOINTS AS DIAGNOSTICS

机译:直接测定循环伏安法的传递系数-以等点诊断。

获取原文
获取原文并翻译 | 示例
       

摘要

Isopoints are observed when cyclic voltammograms recorded at several scan rates are all plotted on the same potential axis, The isopoint is the point where all curves intersect. Here, isopoints are used as a diagnostic under reversible and quasireversible electron transfer conditions, A method is presented for determining the transfer coefficient, alpha, using only the isopoint potential, switching potential, and peak potentials, all readily identifiable points on the voltammograms, It is not necessary to know the diffusion coefficients, standard rate constant, standard potential, electrode area, or solution concentration to apply this method, The isopoint method is applicable for differences in peak potentials of 0.073-0.145 V. The method is verified experimentally by comparing the values of a determined by the isopoint and convolution methods, A linearization of the dimensionless kinetic parameter, psi, developed by Nicholson, with the difference in peak potentials is presented, It is also shown that the average of the peak potentials yields the half-wave potential within 1 mV for differences in the peak potential of less than approximately 0.145 V. Ah models were developed by computer simulation (Digisim).
机译:当以几种扫描速率记录的循环伏安图都绘制在同一电位轴上时,会观察到等值点。等值点是所有曲线相交的点。在这里,等电点被用作可逆和准可逆电子转移条件下的诊断方法,提出了一种仅使用等点电势,开关电势和峰值电势,伏安图上所有易于识别的点来确定传输系数α的方法,不必知道使用该方法的扩散系数,标准速率常数,标准电势,电极面积或溶液浓度,isopoint方法适用于0.073-0.145 V的峰值电势差。该方法通过比较进行了实验验证由等值点法和卷积法确定的a值,给出了由Nicholson开发的无量纲动力学参数psi的线性化,具有峰值电势的差异,并且还显示了峰值电势的平均值产生了一半的1 mV内的最大电势,峰值电势差小于约0.145V。Ah模型s是通过计算机仿真(Digisim)开发的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号