High-Resolution Microcoil NMR for Analysis of Mass-Limited, Nanoliter Samples

机译:高分辨率微线圈NMR用于分析质量受限的纳升样品

获取原文
获取原文并翻译 | 示例
       

摘要

An improved nanoliter-volume NMR probe design places the microcoil and capillary at the magic angle (54.7° with respect to the external magnetic field. Using an NMR probe which requires a total sample volume of just 200 nL, high-resolution 300-MHz 250L?1)H-NMR spectra (line width, 0.6 Hz) are presented of 10 mM α-bag cell peptide for an observe quantity of 45 ng (50 pmol in 5 nL). For the volume of sample inside the microcoil (the observe volume, V↓(obs)), the 3σ limit of detection (LOD) is 9 ng (10 pmol, 2 mM) for data obtained in 15 h. To reduce the data acquisition time, a probe with a greater V↓(obs) is developed. As an example of a rapid, mass-limited analysis, a concentration corresponding to 400 ng of menthol dissolved in V↓(obs) = 31 nL (82.6 mM) yields a spectrum in 9 min (LOD = 6.9 ng, 44 pmol, 1.4 mM). To illustrate improvements in concentration sensitivity, aspectrum is acquired in 45 rain for 400 ng of menthol dissolved in a total sample volume of 200 nL (12.8 mM). Compared to a commercial nanoprobe for the same mass of menthol, these two examples reduce data acquisition time by at least 95%. Both model compounds demonstrate substantially improved concentration LODs compared to those obtained in previous high-resolution, microcoil NMR work. These advances illustrate the utility of enhanced sensitivity provided by NMR microcoils applied to nanoliter volumes of mass-limited samples.
机译:改进的纳升体积NMR探针设计使微线圈和毛细管处于魔角(相对于外部磁场为54.7°)。使用NMR探针仅需200 nL的总样品量,高分辨率300 MHz给出了10 mMα-bag细胞肽的250L?1)H-NMR光谱(线宽,0.6 Hz),观察量为45 ng(5 nL中50 pmol)。对于微线圈内部的样品体积(观察体积,V↓(obs)),对于在15小时内获得的数据,3σ检出限(LOD)为9 ng(10 pmol,2 mM)。为了减少数据采集时间,开发了一种具有更大V↓(obs)的探头。作为快速,有限质量分析的一个示例,相当于400 ng薄荷醇溶解在V↓(obs)= 31 nL(82.6 mM)中的浓度可在9分钟内产生一个光谱(LOD = 6.9 ng,44 pmol,1.4毫米)。为了说明浓度灵敏度的提高,在45雨中获得了400 ng薄荷醇溶解在200 nL(12.8 mM)的总样品量中的长宽比。与相同质量的薄荷醇的商用纳米探针相比,这两个示例将数据采集时间缩短了至少95%。与以前的高分辨率微线圈NMR研究相比,这两种模型化合物均显示出浓度LOD显着提高。这些进展说明了将NMR微线圈应用于纳升体积的质量受限样品所提供的增强灵敏度的实用性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号