The feasibility of quantitative instrumental neutron activation analysis (INAA) of samples in the kilogram range without internal standardization has been demonstrated by Overwater et al. (Anal. Chem. 1996, 68, 341). In their studies, however, they demonstrated only the agreement between the "corrected" γ ray spectrum of homogeneous large samples and that of small samples of the same material. In this paper, the k↓(o) calibration of the IRI facilities for large samples is described, and, this time in terms of (trace) element concentrations, some of Overwater's results for homogeneous materials are presented again, as well as results obtained from inhomogeneous materials and subsamples thereof. It is concluded that large-sample INAA can be as accurate as ordinary INAA, even when applied to inhomogeneous materials.
机译:Overwater等人已经证明了在没有内部标准化的情况下对千克范围内的样品进行定量仪器中子活化分析(INAA)的可行性。 (Anal.Chem.1996,68,341)。但是,在他们的研究中,他们仅证明了均质大样品的“校正”γ射线谱与相同材料的小样品的γ射线谱之间的一致性。本文描述了IRI设施对大样品的k↓(o)校准,并且这次以元素(痕量)的浓度再次介绍了均质材料的Overwater结果,以及获得的结果。从不均匀材料及其子样本中提取。结论是,即使应用于非均质材料,大样本INAA也可以与普通INAA一样准确。