首页> 外文期刊>Biomedical Chromatography: An International Journal Devoted to Research in Chromatographic Methodologies and Their Applications in the Biosciences >Rapid ultraperformance liquid chromatography-tandem mass spectrometry method for quantification of oxcarbazepine and its metabolite in human plasma.
【24h】

Rapid ultraperformance liquid chromatography-tandem mass spectrometry method for quantification of oxcarbazepine and its metabolite in human plasma.

机译:快速超高效液相色谱-串联质谱法定量测定人血浆中的奥卡西平及其代谢产物。

获取原文
获取原文并翻译 | 示例
           

摘要

A rapid and sensitive ultraperformance liquid chromatography tandem mass spectrometry assay was developed for the simultaneous analysis of oxcarbazepine and its main metabolite in human plasma. The assay involves a simple solid-phase extraction procedure of 0.3 mL of human plasma and analysis was performed on a triple-quadrupole tandem mass spectrometer in multiple reaction monitoring mode via electrospray ionization. Separation was achieved on an Acquity UPLC BEH C(18) column (50 x 2.1 mm, i.d., 1.7 microm) with isocratic elution at a flow-rate of 0.25 mL/min and imipramine was used as the internal standard. The standard calibration curve was linear over the range 9.580-5070.205 ng/mL for oxcarbazepine (OXC) and 19.444-10290.800 ng/mL for 10,11-dihydro-10-hydroxycarbamazepine (MHD), expressed by the linear correlation coefficient r(2) , which was better than 0.995 for OXC and MHD. The intra- and inter-day precision and accuracy of the quality control samples were within 10.0%. The recoveries were 81.0, 89.6 and 66.6% for OXC, MHD and imipramine, respectively. The total run time was 1.5 min only for each sample, which makes it possible to analyze more than 350 samples per day. Copyright (c) 2010 John Wiley & Sons, Ltd.
机译:建立了快速灵敏的超高效液相色谱串联质谱分析法,用于同时分析人血浆中的奥卡西平及其主要代谢产物。该测定法涉及0.3 mL人血浆的简单固相萃取程序,并通过电喷雾电离在三反应四极杆串联质谱仪上以多反应监测模式进行了分析。在Acquity UPLC BEH C(18)色谱柱(50 x 2.1 mm,i.d. 1.7微米)上以等度洗脱以0.25 mL / min的流速完成分离,并将丙咪嗪用作内标。奥卡西平(OXC)的标准校准曲线在9.580-5070.205 ng / mL范围内线性,而10,11-二氢-10-羟基卡马西平(MHD)的标准校准曲线范围为19.444-10290.800 ng / mL,由线性相关系数r(2 ),对于OXC和MHD而言,该值优于0.995。质量控制样品的日内和日间精度和准确度在10.0%以内。 OXC,MHD和丙咪嗪的回收率分别为81.0、89.6和66.6%。仅每个样品的总运行时间为1.5分钟,因此每天可以分析350多个样品。版权所有(c)2010 John Wiley&Sons,Ltd.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号