New Polymeric Membrane Electrode for Azithromycin Determination

机译:用于阿奇霉素测定的新型聚合物膜电极

获取原文
获取原文并翻译 | 示例
           

摘要

A new polymeric membrane electrode has been developed for the determination of azithromycin.The electrode was constructed by incorporating the azithromycin-tetraiodomercurate ion pair complex into a polyvinylchloride matrix plasticized by nitrobenzene.This sensor exhibited good linear response (-29.8 mV/ decade) over the concentration range 1.0 x 10~(-2)-7.0 x 10~(-6)M,using a mixed solution as supporting electrolyte containing phosphate buffer (pH = 6.5;I = 0.1 M) and acetonitrile (3:1,v/v).The pH did not affect the electrode performances within the pH range 2-8.Interferences from species such as pharmaceutical excipients are negligible.In order to assess the performance and reliability of the analytical results,the proposed potentiometric method using this selective electrode to azithromycin has been validated.The validated method was successfully applied,using direct potentiometry,to the quantitation of azithromycin in pharmaceutical preparation (rsd=0.87%;mean recovery = 99.5%) and in dissolution profile studies of tablets.
机译:已开发出一种用于测定阿奇霉素的新型聚合物膜电极,该电极是通过将阿奇霉素-四碘草酸酯离子对络合物掺入由硝基苯增塑的聚氯乙烯基质中而构建的,该传感器在整个过程中表现出良好的线性响应(-29.8 mV /十倍)。浓度范围1.0 x 10〜(-2)-7.0 x 10〜(-6)M,使用混合溶液作为支持电解质,包含磷酸盐缓冲液(pH = 6.5; I = 0.1 M)和乙腈(3:1,v / v).pH在2-8的pH范围内不会影响电极性能。药物赋形剂等物质的干扰可忽略不计。为评估分析结果的性能和可靠性,建议使用该选择性电极的电位法直接电位法已成功地应用于药物制剂中阿奇霉素的定量分析(rsd = 0.87%;平均回收率= 99.5%)在片剂的溶出度研究中。