The first sequential injection assay for the generic determination of gabapentin and pregabalin is reported. The analytes react with o-phthalaldehyde in the presence of N-acetylcysteine as a nucleophilic reagent in alkaline medium under flow conditions to form highly fluorescent derivatives. The effect of the main instrumental and chemical variables on the assay was examined. The proposed method was validated for both analytes in terms of linearity, detection, and quantitation limits (cL=160 lgL-1, cQ=480 lgL-1 for gabapentin, and cL=70 lgL-1, cQ=210 lgL-1 for pregabalin), precision (sr<1.0% in both cases), selectivity, and accuracy. The applicability of the assay was demonstrated by successfully analyzing commercially available formulations. The experimental percent recoveries were in the range of 97.9–102.0% for gabapentin and 98.3–102.3% for pregabalin.
机译:报道了用于通用测定加巴喷丁和普瑞巴林的首次顺序注射测定法。在碱性条件下,流动条件下,分析物与邻苯二甲醛在作为亲核试剂的N-乙酰半胱氨酸存在下反应,形成高度荧光的衍生物。检查了主要仪器和化学变量对测定的影响。在线性,检测和定量限方面对两种分析物都验证了所提出的方法(加巴喷丁的cL = 160 lgL-1,cQ = 480 lgL-1,cL = 70 lgL-1,cQ = 210 lgL-1普瑞巴林),精密度(两种情况下sr <1.0%),选择性和准确度。通过成功分析市售制剂,证明了该测定法的适用性。加巴喷丁的实验回收率在97.9-102.0%范围内,普瑞巴林的回收率在98.3-102.3%范围内。