Verification Study of Residual Activity Measurements After Yttrium-90 Radioembolization with Glass Microspheres

机译:用玻璃微球钇-90放射栓塞后残留活性测量的验证研究

获取原文
获取原文并翻译 | 示例
       

摘要

Objective After yttrium-90 (Y-90) radioembolization, residual activity and its consequences for dosimetric calculations are often not reported. The manufacturer for glass microspheres prescribes standard residual activity measurements by a survey meter, but the validity lacks evidence. This study aims to verify the accuracy of the survey meter approach for measuring residual activity of glass microspheres after treatment with glass microspheres. Methods To validate the accuracy of the survey meter approach, the measured residual activity of glass microspheres by survey meter was compared with measurements by PET. A sample of these waste containers was also measured by dose calibrator to confirm the accuracy of the PET. Results Twenty-four waste containers from glass microsphere treatments were prospectively scanned with Y-90-PET/CT. Bland-Altman plots showed substantial disagreement in residual activity measured by survey meter versus the residual activity measured by PET and dose calibrator, whereas the correlation between PET and dose calibrator was excellent (rho = 0.99). Conclusion This study found a significant disagreement between the residual activities measured by the survey meter, compared to measurements by PET and dose calibrator. If relatively high amounts of residual activity are encountered using the exposure rate measurement with a survey meter, additional quantification should be considered using either PET/CT or a dose calibrator measurement.
机译:目的钇-90(Y-90)放射性栓塞后,剩余活性及其对剂量学计算的影响常常没有报道。玻璃微球制造商规定了用测量仪进行标准剩余活度测量,但其有效性缺乏证据。本研究旨在验证测量仪法测量玻璃微球治疗后剩余活性的准确性。方法将测量仪测量的玻璃微球剩余活性与PET测量结果进行比较,验证测量仪方法的准确性。这些废物容器的样本也通过剂量校准器进行测量,以确认PET的准确性。结果用Y-90-PET/CT对24个玻璃微球处理后的废物容器进行了前瞻性扫描。Bland Altman图显示,测量仪测得的剩余活度与PET和剂量校准器测得的剩余活度存在显著差异,而PET和剂量校准器之间的相关性非常好(rho=0.99)。结论本研究发现,与PET和剂量校准器测量相比,测量仪测量的残余活度存在显著差异。如果使用测量仪的暴露率测量遇到相对较高的残余放射性,则应考虑使用PET/CT或剂量校准器测量进行额外量化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号