首页> 外文期刊>Analytical chemistry >A Multicomponent Mobile Phase for Ion Chromatography Applied to the Separation of Anions from the Residue of Low Explosives
【24h】

A Multicomponent Mobile Phase for Ion Chromatography Applied to the Separation of Anions from the Residue of Low Explosives

机译:一种用于离子色谱的多组分流动相,用于从低炸药残留物中分离阴离子

获取原文
获取原文并翻译 | 示例
       

摘要

A multicomponent mobile phase utilizing ion-exchange, ion-exclusion, and ion-pairing principles for the rapid isocratic separation of anions in low explosives residue by ion chromatography (IC) has been developed. The notable feature of this system is that an ion-pairing reagent and an ion-exclusion reagent are combined in the same mobile phase. Contrary to expectation, these reagents act independently of each other in solution. The stock mobile-phase composition consisted of boric acid, D-gluconic acid, lithium hydroxide, and glycerol. Tetrapropylammonium hydroxide, an ion-interaction reagent was used to achieve pH 8.5. Acetonitrile (ACN) was added to enhance resolution and octanesulfonic acid, an ion-exclusion reagent, was added to adjust the retention time of perchlorate. Separation of a mixture of anions common to low explosives residue was achieved in less than 16 min using a Waters IC-Pak Anion HR column. Optimization studies were performed by changing the concentration of the ACN and by altering the pH or the type of ion-interaction or -exclusion agents. Simulated case studies were performed using postblast residues from pipe bombs. The results show this method to be a valid and reproducible procedure for forensic casework analysis. The practical significance of this system is that a reduction in the analysis time and an improvement in efficiency of late-eluting peaks can be achieved without resorting to gradient elution techniques. For the analysis of anions detected in explosives residue, the Waters IC-Pak Anion HR column has proven to be a suitable replacement for the Vydac 300IC405 column, which has been discontinued by the manufacturer.
机译:已开发出一种利用离子交换,离子排斥和离子对原理的多组分流动相,用于通过离子色谱法(IC)快速等度分离低炸药残留物中的阴离子。该系统的显着特点是离子配对试剂和离子排斥试剂在同一流动相中混合使用。与预期相反,这些试剂在溶液中彼此独立地起作用。储备流动相组合物由硼酸,D-葡萄糖酸,氢氧化锂和甘油组成。使用四丙基氢氧化铵(一种离子相互作用试剂)达到pH 8.5。加入乙腈(ACN)以提高分离度,并加入辛烷磺酸(一种离子排斥试剂)来调节高氯酸盐的保留时间。使用Waters IC-Pak Anion HR色谱柱在不到16分钟的时间内即可分离出易爆炸药残留的常见阴离子混合物。通过改变ACN的浓度并改变pH或离子相互作用或排斥剂的类型来进行优化研究。使用来自管炸弹的爆炸后残留物进行了模拟案例研究。结果表明,该方法是法医案件分析的有效且可重复的过程。该系统的实际意义在于,无需诉诸梯度洗脱技术,就可以减少分析时间并提高后洗脱峰的效率。为了分析炸药残留物中检测到的阴离子,沃特世IC-Pak阴离子HR色谱柱已被证明是Vydac 300IC405色谱柱的合适替代品,该色谱柱已被制造商停产。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号