A new electrochemical biosensing strategy has been developed for trace measurements of toxic aromatic amine compounds. The device relies on the intercalative collection of aromatic amines onto the immobilized dsDNA layer followed by potentiometric stripping quantitation of the accumulated species. The enhanced sensitivity, accrued from the DNA collection process, is coupled to new selectivity dimensions provided by the structural requirements for such intercalative binding. The extent and rate of theaccumulation are strongly dependent upon the structure of the aromatic amine species. Having the amino substituent in a slightly different position produces a dramatic effect upon the response. Nanomolar detection limits are obtained after a 10-min accumulation. Applicability to river water and groundwater samples is demonstrated. Such DNA-based devices hold great promise for environmental screening of toxic aromatic amines and for elucidating molecular interactions between intercalating pollutants with DNA.
机译:已经开发出一种新的电化学生物传感策略,用于痕量有毒芳香胺化合物的测量。该设备依赖于芳香胺的插层收集到固定化的dsDNA层上,然后对累积的物质进行电位溶出定量分析。从DNA收集过程中获得的增强的灵敏度与这种插入结合的结构要求所提供的新的选择性维度相关。积累的程度和速率在很大程度上取决于芳族胺类的结构。氨基取代基的位置略有不同会对响应产生显着影响。累积10分钟后获得纳摩尔检测限。证明了对河水和地下水样品的适用性。这种基于DNA的设备在环境筛选有毒的芳香胺和阐明嵌入的污染物与DNA之间的分子相互作用方面具有广阔的前景。