A simple way to selectively isolate peptides based on their isoelectric points(pi)for MALDI mass spectral analysis is described.An applied voltage was used to electromi-grate peptides into a capillary.The capillary was modified with a zwitterionic surfactant,1,2-dilauroyl-sre-phosphati-dylcholine(DLPC),to suppress the electroosmotic flow(EOF)during injection.Hence,eitiier the cationic or the anionic peptides were introduced,depending on the voltage polarity.By controlling the pH,selective loading of peptides was performed to isolate trace components from a mixture.The injected sample plugs were subse-quendy spotted in nanoliter volumes for MALDI-MS analysis.No significant sample losses resulting from selective sampling were detected.Low attomole-level detection of peptides(adrenocorticotropic hormone fragment 18-39,pi 4.25)was achieved from a mixture containing other peptides(angiotensin I,pi 6.92,and bradykinin,pi 12.00)at 100 000-fold higher concentrations.
机译:描述了一种基于等电点(pi)选择性分离肽进行MALDI质谱分析的简单方法。使用施加的电压将肽电迁移到毛细管中。用两性离子表面活性剂1,2-修饰毛细管二月桂酰-sre-磷脂酰胆碱(DLPC),以抑制注射过程中的电渗流(EOF)。因此,根据电压极性引入阳离子或阴离子肽,通过控制pH值,可选择性地加载肽进行了分离以从混合物中分离出痕量成分。将注入的样品塞子以亚升体积次次斑点定位以进行MALDI-MS分析。没有发现选择性采样导致的显着样品损失。低原子水平的肽检测(促肾上腺皮质激素片段18 -39,pi 4.25)是由含有其他肽(血管紧张素I,pi 6.92和缓激肽,pi 12.00)的混合物以高100000倍的浓度获得的。