首页> 外文期刊>Analytical chemistry >Separation of cisplatin and its hydrolysis products using electrospray ionization high-field asymmetric waveform ion mobility spectrometry coupled with ion trap mass spectrometry
【24h】

Separation of cisplatin and its hydrolysis products using electrospray ionization high-field asymmetric waveform ion mobility spectrometry coupled with ion trap mass spectrometry

机译:电喷雾电离高场非对称波形离子迁移谱法与离子阱质谱联用分离顺铂及其水解产物

获取原文
获取原文并翻译 | 示例
       

摘要

Cisplatin and its mono- and dihydrated complexes have been separated using a high-field asymmetric waveform ion mobility spectrometry (FAIMS) analyzer interfaced with electrospray ionization (ESI) and ion trap mass spectrometry (ITMS). The addition of helium to the nitrogen curtain/carrier gas in the FAIMS device improved both the sensitivity and selectivity of the electrospray analysis. Introduction of a three-component mixture as curtain/carrier gas, nitrogen, helium, and carbon dioxide, resulted in further improvements to sensitivity. Compared with conventional ESI-MS, the background chemical noise in the ESI-FAIMS-ITMS spectrum was dramatically reduced, resulting in over 30-fold improvement in the signal-to-noise ratio for cisplatin. Analytical results were linear over the concentration range 10-200 ng/mL for intact cisplatin with a corresponding detection limit determined of 0.7 ng/mL with no derivatization or chromatographic separation prior to analysis. [References: 37]
机译:使用与电喷雾电离(ESI)和离子阱质谱(ITMS)连接的高场非对称波形离子迁移谱(FAIMS)分析仪已分离了顺铂及其一水合物和二水合物。在FAIMS设备中,向氮气幕/载气中添加氦气可以提高电喷雾分析的灵敏度和选择性。引入三组分混合物作为幕布/载气,氮气,氦气和二氧化碳,进一步提高了灵敏度。与传统的ESI-MS相比,ESI-FAIMS-ITMS光谱中的背景化学噪声大大降低,从而使顺铂的信噪比提高了30倍以上。对于完整的顺铂,分析结果在10-200 ng / mL的浓度范围内是线性的,相应的检测限为0.7 ng / mL,分析前没有衍生化或色谱分离。 [参考:37]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号