首页> 外文期刊>Analytical chemistry >Zeptomole-sensitivity electrospray ionization - Fourier transform ion cyclotron resonance mass spectrometry of proteins
【24h】

Zeptomole-sensitivity electrospray ionization - Fourier transform ion cyclotron resonance mass spectrometry of proteins

机译:Zeptomole-敏感性电喷雾电离-傅立叶变换离子回旋共振蛋白质质谱

获取原文
获取原文并翻译 | 示例
       

摘要

Methods are being developed for ultrasensitive protein characterization based upon electrospray ionization (ESI) with Fourier transform ion cyclotron resonance mass spectrometry (FTICR-MS). The sensitivity of a FTICR mass spectrometer equipped with an ESI source depends on the overall ion transmission, which combines the probability of ionization, transmission efficiency, and ion trapping in the FTICR cell. Our developments implemented in a 3.5 tesla FTICR mass spectrometer include introduction and optimization of a newly designed electrodynamic ion funnel in the ESI interface, improving the ion beam characteristics in a quadrupole-electrostatic ion guide interface, and modification of the electrostatic ion guide. These developments provide a detection limit of approximately 30 zmol (similar to 18 000 molecules) for proteins with molecular weights ranging from 8 to 20 kDa. [References: 23]
机译:正在开发基于电喷雾电离(ESI)和傅立叶变换离子回旋共振质谱(FTICR-MS)的超灵敏蛋白质表征方法。配备ESI源的FTICR质谱仪的灵敏度取决于整个离子传输,该离子传输结合了电离的概率,传输效率和FTICR池中的离子捕获。我们在3.5特斯拉FTICR质谱仪中实现的开发包括在ESI界面中引入和优化新设计的电动离子漏斗,改善四极杆-静电离子导向器界面中的离子束特性以及对静电离子导向器进行修改。这些发展为分子量为8至20 kDa的蛋白质提供了约30 zmol(类似于18 000分子)的检测极限。 [参考:23]