A fiber-optic sensor for the continuous and simultaneous determination of glucose and oxygen is described. The sensor is comprised of dual-analyte sensing sites in defined positions on the distal end of an imaging fiber (350 mu m o.d.). Each sensing site is an individual polymer cone covalently attached to the activated fiber surface using localized photopolymerization. The oxygen sensor consists of a double-layer polymer cone. The inner polymer cone is a hydrophobic gas-permeable copolymer containing an oxygen-sensitive ruthenium dye, and the outer layer is a poly(hydroxyethyl methacrylate) (HEMA) polymer. The glucose sensor is an oxygen sensor with a poly-HEMA outer layer containing immobilized glucose oxidase. The fluorescence images of both sensing sites are captured with a CCD camera, and the measured fluorescence intensities are related to analyte concentrations. Oxygen quenching data for both sensing sites fit a two-site Stern-Volmer quenching model. The sensor has been used to simultaneously monitor independent changes in glucose and oxygen concentrations. Glucose calibration curves were obtained under varying oxygen tensions, and the detection limit is 0.6 mM glucose. The effect of fluctuations in oxygen partial pressure on the glucose response can be used to calibrate the sensor. The sensor response time varies from 9 to 28 s, depending on the different thicknesses of the enzyme layer. The sensor maintains the same sensitivity for 2 days. Multiple glucose sensing sites with different enzymatic activities can be immobilized on the distal end of the fiber, affording control of the linear range.
机译:描述了一种用于连续和同时确定葡萄糖和氧气的光纤传感器。该传感器由位于成像光纤远端(350微米)的确定位置的双分析物传感位点组成。每个感测位点是使用局部光聚合共价附着到活化纤维表面的单个聚合物锥体。氧气传感器由双层聚合物锥体组成。内聚合物锥体是包含对氧敏感的钌染料的疏水性可透气的共聚物,外层是聚甲基丙烯酸羟乙酯(HEMA)聚合物。葡萄糖传感器是具有包含固定的葡萄糖氧化酶的聚HEMA外层的氧气传感器。两个感测部位的荧光图像均由CCD相机捕获,并且测得的荧光强度与分析物浓度相关。两个感测部位的氧气淬火数据均符合两部位Stern-Volmer淬火模型。该传感器已用于同时监测葡萄糖和氧气浓度的独立变化。在不同的氧气张力下获得了葡萄糖校准曲线,检出限为0.6 mM葡萄糖。氧分压波动对葡萄糖反应的影响可用于校准传感器。传感器响应时间从9到28 s不等,具体取决于酶层的厚度。传感器保持相同的灵敏度2天。可以将具有不同酶活性的多个葡萄糖感测位点固定在纤维的远端,从而控制线性范围。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号