首页> 外文期刊>Analytical chemistry >Liquid-liquid extraction in the 96-well plate format with SRM LC MS quantitative determination of methotrexate and its major metabolite in human plasma
【24h】

Liquid-liquid extraction in the 96-well plate format with SRM LC MS quantitative determination of methotrexate and its major metabolite in human plasma

机译:用SRM LC MS定量测定96孔板形式的液-液定量测定人血浆中的甲氨蝶呤及其主要代谢产物

获取原文
获取原文并翻译 | 示例
           

摘要

A method involving the semirobotic liquid-liquid extraction (LLE) in deep-well 96-well plates was developed for the quantitation of the anti-cancer/antiinflammatory drug methotrexate (MTX) and its major metabolite, 7-hydroymethotrexate (70H-MTX) in human plasma. The extraction time for the sample preparation was relatively short with four 96-well plates (384 samples) prepared in approximately 90 min by one person. The sample extracts were each analyzed within 1.2 min using a positive ion turbo-ionspray selected reaction monitoring liquid chromatography/mass spectrometric (SRM LC/MS) method in which 768 samples were easily analyzed within 22 h (maximum of 820 samples in 24 h). Deuterated internal standards, MTX-d(3) and 70H-MTX-d(3), were used. The calibration curves for MTX and 70H-MTX were linear (R-2 > 0.997) and ranged from 0.5 to 250 and 0.75 to 100 ng/mL, respectively. The limit of quantitation (LOQ) for MTX and 70H-MTX was 0.5 and 0.75 ng/mL, respectively; persistent carryover from the autosampler limited the LOQ achievable. The limit of detection (LOD) was 0.05 ng/mL for MTX and 0.1 ng/mL for 70H-MTX The intra- and inter-assay precision and accuracy did not exceed 15% for both MTX and 70H-MTX. The recoveries were 61% for MTX and 47% for 70H-MTX The method was validated and demonstrated to be robust with high precision and accuracy. [References: 26]
机译:开发了一种涉及在深孔96孔板中进行半机器人液液萃取(LLE)的方法,用于定量抗癌/抗炎药甲氨蝶呤(MTX)及其主要代谢物7-羟甲氨蝶呤(70H-MTX)在人类血浆中。用一个人在大约90分钟内制备的四个96孔板(384个样品)中,样品制备的提取时间相对较短。使用正离子涡轮离子喷雾选择的反应监测液相色谱/质谱(SRM LC / MS)方法在1.2分钟内分析样品提取物,其中22小时内可轻松分析768个样品(24小时内最多分析820个样品) 。使用氘代内标MTX-d(3)和70H-MTX-d(3)。 MTX和70H-MTX的校准曲线是线性的(R-2> 0.997),范围分别为0.5至250 ng / mL和0.75至100 ng / mL。 MTX和70H-MTX的定量限(LOQ)分别为0.5和0.75 ng / mL。自动进样器的持续残留限制了可实现的最低定量。 MTX的检出限(LOD)为0.05 ng / mL,70H-MTX的检出限为0.1 ng /mL。MTX和70H-MTX的批内和批间精密度和准确性均不超过15%。 MTX的回收率为61%,70H-MTX的回收率为47%。该方法已经过验证并证明是可靠的,具有高精度和高精度。 [参考:26]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号