A small-volume heterogeneous immunoassay system is demonstrated in microchannels exploiting magnetic manipulations of small paramagnetic particles (1-2-mum diameter). The small-diameter particles help to create a high surface-area-to-volume ratio that generates the sensitivity for the small detection volumes. Flow characteristics of the magnetically formed packed bed within the channel allow the assay to be carried out quickly (minutes) while passing appropriate volumes of both samples and reagents (microliters to hundreds of nanoliters) through the system. The assay is demonstrated with a direct interaction of fluorescein isothiocyanate (FITC) with an immobilized anti-FITC conjugate in which a small-volume sample (< 1 muL) reaches 90% of maximum signal in 3 min. Heterogeneous sandwich assays are demonstrated with parathyroid hormone (PTH) and interleukin-5 (IL-5). Both the PTH assay and IL-5 assays were carried out on microliter volumes and demonstrated physiologically relevant sensitivity (similar to mug/L).
机译:利用微顺磁性小颗粒(1-2微米直径)的磁操作,在微通道中证明了小体积异质免疫分析系统。小直径的颗粒有助于产生高的表面积体积比,从而为小检测体积产生灵敏度。通道内磁性形成的填充床的流动特性使分析能够快速(几分钟)进行,同时使适当体积的样品和试剂(微升至数百纳升)通过系统。荧光素异硫氰酸酯(FITC)与固定的抗FITC偶联物直接相互作用证明了该测定法,其中小体积样品(<1μL)在3分钟内达到最大信号的90%。甲状旁腺激素(PTH)和白介素5(IL-5)证实了异质夹心测定。 PTH测定法和IL-5测定法均以微升体积进行,并显示出生理相关的敏感性(类似于杯子/升)。