首页> 外文期刊>Analytical chemistry >Mass Transfer and Gas-Phase Calibration of Implanted Oxygen Sensors
【24h】

Mass Transfer and Gas-Phase Calibration of Implanted Oxygen Sensors

机译:植入式氧气传感器的传质和气相校准

获取原文
获取原文并翻译 | 示例
           

摘要

A protocol is described for validation of implanted oxygen sensors, in which sensors are calibrated in the gas phase where concentration boundary layers are absent. Calibration prior to sensor implantation and confirmation after sensor explantation allows separation of tissue mass transfer effects from sensor variance and drift. A model is given here that describes the oxygen-dependent signal current in terms of oxygen mass transfer to the sensor, permeability of the sensor membrane, and electrode area. The parameter used in the model to describe mass transfer to implanted sensors is consistent with experimental observations and allows comparisons with non-implanted sensors. This method provides a bridge between the complementary approaches of empirical calibration and model-based calculation for determining oxygen concentration from the sensor response.
机译:描述了用于验证所植入的氧气传感器的协议,其中在没有浓度边界层的气相中校准传感器。传感器植入之前的校准和传感器植入之后的确认允许将组织传质效应与传感器差异和漂移区分开。此处给出一个模型,该模型根据氧气向传感器的质量转移,传感器膜的渗透性和电极面积来描述与氧气有关的信号电流。模型中用于描述向植入传感器的质量转移的参数与实验观察结果一致,并且可以与非植入传感器进行比较。该方法为经验校准和基于模型的计算的补充方法之间的桥梁,用于根据传感器响应确定氧气浓度。