摘要

A capillary isoelectric focusing (CIEF) system with liquid core waveguide (LCW) laser-induced fluorescence whole column imaging detection was developed in this study. A Teflon AF 2400 capillary was used as both the separation channel and the axially illuminated LCW. The excitation light was introduced at one end of the capillary, and propagated forward within the capillary. As the Teflon AF 2400 capillary has a refractive index (n = 1.29-1.31) lower than that of water (n = 1.33), total internal reflection was very apparent. The employment of the Teflon AF 2400 capillary avoided the use of high refractive index additives such as glycerol, accommodating the system to wider applications. Due to its inert chemical properties, the capillary exhibited limited protein adsorption and electroosmotic flow; thus, the need for capillary preconditioning with polymeric solution and the addition of polymeric additives into the sample mixture can be avoided. Three types of proteins, naturally fluorescent proteins, covalently labeled proteins, and noncovalently labeled proteins, were examined using this method. CIEF under denaturing conditions was also explored, and several advantages over the native mode were found. When compared to a commercially available instrument with UV detection, the separation efficiency and peak capacity were similar while the detection sensitivity was enhanced by 3-5 orders of magnitude. [References: 45]
机译:本研究开发了一种带有液芯波导(LCW)激光诱导的荧光全柱成像检测的毛细管等电聚焦(CIEF)系统。将Teflon AF 2400毛细管用作分离通道和轴向照亮的LCW。激发光在毛细管的一端引入,并在毛细管内向前传播。由于Teflon AF 2400毛细管的折射率(n = 1.29-1.31)低于水的折射率(n = 1.33),因此全内反射非常明显。铁氟龙AF 2400毛细管的使用避免了使用高折射率的添加剂(如甘油),从而使系统适应更广泛的应用。由于其惰性的化学特性,毛细管显示出有限的蛋白质吸附和电渗流;因此,可以避免使用聚合物溶液进行毛细管预处理以及将聚合物添加剂添加到样品混合物中的需求。使用此方法检查了三种类型的蛋白质,即天然荧光蛋白质,共价标记的蛋白质和非共价标记的蛋白质。还探讨了变性条件下的CIEF,并发现了优于纯模式的几个优势。与市售的带紫外线检测的仪器相比,分离效率和峰容量相似,而检测灵敏度提高了3-5个数量级。 [参考:45]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号