Detection limit of isotope dilution mass spectrometry

机译:同位素稀释质谱法的检出限

获取原文
获取原文并翻译 | 示例
           

摘要

The detection limit is an important figure of merit for evaluating instrumentation and analytical methods. While the detection limit for techniques using linear calibration functions has been studied extensively, this fundamental metric has rarely been discussed for mass spectrometry that bases the calibration on the principle of isotope dilution.,We have developed a formulation for the detection-limit for isotope dilution mass spectrometry (IDMS) after a thorough analysis of the uncertainty of IDMS measurements. The new formulation describes the, IDMS detection limit as a function of the enrichment of the isotopic spike and the linear calibration detection limits measured at the masses for the isotope ratio measurement. [References: 8]
机译:检出限是评估仪器和分析方法的重要指标。尽管已经广泛研究了使用线性校准功能的技术的检测限,但基于同位素稀释原理进行校准的质谱法很少讨论该基本指标。我们已经开发了同位素稀释检测限的配方彻底分析IDMS测量的不确定性后进行质谱分析(IDMS)。新的配方将IDMS的检出限描述为同位素峰的富集的函数,并以同位素比测量的质量测量了线性校准检出限。 [参考:8]