首页> 外文期刊>Analytical chemistry >Needle-type dual microsensor for the simultaneous monitoring of glucoseand insulin
【24h】

Needle-type dual microsensor for the simultaneous monitoring of glucoseand insulin

机译:针型双重微传感器,可同时监测葡萄糖和胰岛素

获取原文
获取原文并翻译 | 示例
       

摘要

A miniature needle-type sensor suitable for the simultaneous amperometric monitoring of glucose and insulin is described. The integrated microsensor consists of dual (biologically and chemically) modified carbon-paste working electrodes inserted into a 14-guage needle. The glucose probe is based on the biocatalytic action of glucose oxidase, and the insulin one relies on the electrocatalytic activity of ruthenium oxide. The analytical performance of the dual sensor is assessed under flow injection conditions, The needle dual detector exhibits a very rapid response to dynamic changes in the concentrations of glucose and insulin. No apparent cross reactivity is observed in mixtures containing millimolar glucose levels and nanomolar insulin concentrations, The response is highly linear (to at least 1000 nM insulin and 14 mM glucose) and reproducible (RSD = 2.6-4.1%). The combination microsensor holds great promise for real-time measurements of the insulin/glucose ratio and for improved management of diabetes.
机译:描述了适用于同时进行葡萄糖和胰岛素的安培监测的微型针型传感器。集成式微传感器由双(生物和化学)修饰的碳糊工作电极组成,该电极插入14针的针中。葡萄糖探针基于葡萄糖氧化酶的生物催化作用,而胰岛素则依赖于氧化钌的电催化活性。在流动注射条件下评估双传感器的分析性能。针双检测器对葡萄糖和胰岛素浓度的动态变化表现出非常快速的响应。在包含毫摩尔葡萄糖水平和纳摩尔胰岛素浓度的混合物中未观察到明显的交叉反应性。该反应是高度线性的(对至少1000 nM胰岛素和14 mM葡萄糖)并且可重复(RSD = 2.6-4.1%)。组合式微传感器在实时测量胰岛素/葡萄糖比和改善糖尿病管理方面具有广阔的前景。