首页> 外文期刊>Applied optics >DMD-based three-dimensional chromatic confocal microscopy
【24h】

DMD-based three-dimensional chromatic confocal microscopy

机译:基于DMD的三维色度共聚焦显微镜

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, a digital mirror device (DMD)-based chromatic confocal microscopy is proposed and demonstrated for three-dimensional (3D) surface profiling without any mechanical scanning. In this method, the DMD works as the multipoint source and multi-pinhole at the same time to achieve the lateral scanning. Moreover, axial scanning is realized through the chromatic aberration of the confocal optics. Since the micromirror array of the DMD is not perpendicular to the confocal imaging axis, a corresponding calibration is needed to eliminate the tilt effects and perform accurate 3D imaging. The measurement range with the current optical system is 45 mu m over 505-650 nm working spectrum and can be increased by using a custom objective with large chromatic aberration. The system performance has been demonstrated with a multistep sample. (C) 2020 Optical Society of America
机译:本文提出了一种基于数字镜装置(DMD)的彩色共焦显微镜,并对无任何机械扫描的三维(3D)表面分析进行了证明。 在此方法中,DMD同时用作多点源和多针孔以实现横向扫描。 此外,通过共聚焦光学器件的色差实现轴向扫描。 由于DMD的微镜阵列不垂直于共聚焦成像轴,因此需要相应的校准来消除倾斜效果并执行精确的3D成像。 具有电流光学系统的测量范围为45μm超过505-650nm的工作频谱,并且可以通过使用大色差的定制物镜来增加。 系统性能已通过多步骤示例进行了演示。 (c)2020美国光学学会