首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Feasibility of the use of a pixelated transmission chamber to measure scattered-radiation in projection radiography: I. Monte Carlo analysis of x-ray scatter propagation and detector modeling
【24h】

Feasibility of the use of a pixelated transmission chamber to measure scattered-radiation in projection radiography: I. Monte Carlo analysis of x-ray scatter propagation and detector modeling

机译:使用像素化传输室的可行性以测量投影射线照相中的散射辐射:I. X射线散射传播和探测器造型的蒙特卡罗分析

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, we propose the use of a pixelated transmission chamber, placed between the patient and the imaging detector, to measure the scatter component of a radiation beam impinging on said imaging detector. Using Monte Carlo simulation, a three-parameter model for the propagation of the scatter compbnent in the transmitted beam is first developed. The use of the transmission chamber to determine the model parameters is then modeled, again using Monte Carlo simulation, and the feasibility of this approach is determined. The amount of radiation backscattered from the imaging detector into the transmission chamber was alsb calculated, for several separation distances between the two. It is shown that at a separation of 10 cm, the amount of backscatter radiation is independent of the imaged object and that therefore it can be determined as part of a calibration procedure for the transmission chamber.
机译:在本文中,我们提出了在患者和成像检测器之间放置的像素化传输室,以测量撞击在所述成像检测器上的辐射束的散射分量。 使用Monte Carlo仿真,首先开发出传输光束中的散射组合在传输的散射组合的三参数模型。 然后使用蒙特卡罗模拟再次建模透射室以确定模型参数,并确定该方法的可行性。 从成像检测器到透射室中的辐射量被计算为ALSB,对于两者之间的几个分离距离。 结果表明,在10cm的分离时,反向散射辐射的量与成像对象无关,因此可以确定作为传输室的校准过程的一部分。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号