Measurements of air kerma index in computed tomography: A comparison among methodologies

机译:计算断层扫描中空气kerma指标的测量:方法中的比较

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

As CT exams impart high doses to patients in comparison to other radiologist techniques, reliable dosimetry is required. In this work, dosimetry in CT beams was carried out in terms of air kerma index in air or in a phantom measured by a pencil ionization chamber, thermoluminescent (TL) detectors and radiochromic film. Calibration results showed the low energy dependence of all three dosimetric systems for the 100-120 kV range, the very high uncertainty of the TL dosimeters in comparison to the other systems and high nonlinearity response in terms of air kerma of the radiochromic film. Measurements with the three systems in a 120 kV CT protocol showed an acceptable agreement among the weighted air kerma index values, but TL dosimeters presented the highest uncertainties in the values.
机译:随着与其他放射科学技术相比,CT考试赋予患者高剂量,需要可靠的剂量测定。 在这项工作中,CT光束的剂量法在空气中的空气中或通过铅笔电离室,热敏发光(TL)检测器和放射刻度膜测量的虚线来进行。 校准结果表明,对于100-120 kV范围的所有三个剂量系统的低能量依赖性,与其他系统和高非线性响应在放射刻度膜的空气Kerma方面相比,TL剂量计的非常高的不确定性。 在120kV CT协议中的三个系统的测量显示了加权空气Kerma指数值之间的可接受协议,但TL剂量计呈现了值的最高不确定性。