首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Results of the CCRI(II)-S12.H-3 supplementary comparison: Comparison of methods for the calculation of the activity and standard uncertainty of a tritiated-water source measured using the LSC-TDCR method
【24h】

Results of the CCRI(II)-S12.H-3 supplementary comparison: Comparison of methods for the calculation of the activity and standard uncertainty of a tritiated-water source measured using the LSC-TDCR method

机译:CCRI(II)-S12.H-3补充比较:使用LSC-TDCR方法测量的氚化水源的活性和标准不确定性的方法比较

获取原文
获取原文并翻译 | 示例
           

摘要

A comparison of calculations of the activity of a (H2O)-H-3 liquid scintillation source using the same experimental data set collected at the LNE-LNHB with a triple-to-double coincidence ratio (TDCR) counter was completed. A total of 17 laboratories calculated the activity and standard uncertainty of the LS source using the files with experimental data provided by the LNE-LNHB. The results as well as relevant information on the computation techniques are presented and analysed in this paper. All results are compatible, even if there is a significant dispersion between the reported uncertainties. An output of this comparison is the estimation of the dispersion of TDCR measurement results when measurement conditions are well defined.
机译:完成了使用在LNE-LNHB上收集的相同实验数据的液体闪烁源的测量的比较,采用在LNE-LNHB上具有三维双重合比(TDCR)计数器。 共有17个实验室计算LS源的活动和标准不确定性使用LNE-LNHB提供的实验数据的文件。 本文提出和分析了关于计算技术的结果以及相关信息。 所有结果都兼容,即使报告的不确定性之间存在显着的分散。 该比较的输出是估计TDCR测量结果的分散,当测量条件明确定义时。