Applying a low energy HPGe detector gamma ray spectrometric technique for the evaluation of Pu/Am ratio in biological samples

机译:应用低能量HPGE探测器伽马射线光谱技术在生物样品中对PU / AM比的评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The estimation of Pu/Am-241 ratio in the biological samples is an important input for the assessment of internal dose received by the workers. The radiochemical separation of Pu isotopes and Am-241 in a sample followed by alpha spectrometry is a widely used technique for the determination of Pu/Am-241 ratio. However, this method is time consuming and many times quick estimation is required. In this work, Pu/Am-241 ratio in the biological sample was estimated with HPGe detector based measurements using gamma/X-rays emitted by these radionuclides. These results were compared with those obtained from alpha spectroscopy of sample after radiochemical analysis and found to be in good agreement. (C) 2015 Elsevier Ltd. All rights reserved.
机译:生物样品中PU / AM-241比的估计是评估工人收到的内部剂量的重要意义。 PU同位素和AM-241在样品中的放射化学分离,其次是α光谱法的测定方法是众所周受的技术,用于测定PU / AM-241比率。 但是,这种方法是耗时,并且需要许多次数快速估计。 在这项工作中,使用这些放射性核素发出的γ/ X射线估计生物样品中的PU / am-241比率基于HPGE检测器的测量。 将这些结果与放射性化学分析之后从样品的α光谱法获得的结果进行了比较,并达成了良好的一致性。 (c)2015 Elsevier Ltd.保留所有权利。