Significant Reduction of Analysis Time with Bulk Sensors Operating in Nonequilibrium Mode

机译:通过以非质量模式运行的批量传感器显着降低分析时间

获取原文
获取原文并翻译 | 示例
       

摘要

The concept of reducing the analysis time with the bulk sensors operating in nonequilibrium mode is formulated theoretically and verified experimentally with ion-selective polymeric optodes. The influence of the measurement time on the optical response range, span, median, and sensitivity is calculated and demonstrated experimentally. Model aqueous samples are successfully analyzed using a nonequilibrium measurement protocol with conventional sodium-selective optodes and with pH optodes containing lipophilic organic electrolyte. The analysis time is significantly reduced from 300 to ca. 100 s, with sufficient recovery, reproducibility, and precision of the sample determination.
机译:通过在非纤基模式下操作的散装传感器减少分析时间的概念理论上并通过离子选择性聚合物光学实验进行实验验证。 实验计算和演示测量时间对光学响应范围,跨度,中值和灵敏度的影响。 使用非常规钠选择性光电和含有亲脂有机电解质的pH光电池成功分析了模型水性样品。 分析时间从300到CA显着降低。 100 S,具有足够的恢复,再现性和样品测定的精度。