Direct Photothermal Measurement of Optical Absorption in a Flow System

机译:流动系统中光学吸收的直接光热测量

获取原文
获取原文并翻译 | 示例
           

摘要

We describe here a simple photothermal detection scheme in a flow stream based on the temperature difference upstream and downstream of the point of illumination. We use a single, two-junction 25 mu m diameter thermocouple to measure the temperature change. The baseline standard deviation in the dark is similar to 0.001 degrees C that increases up to 0.0016 degrees C depending on the illumination source. We demonstrate the detection of several chromatographically separated dyes both with a 1.5 mm and a 0.1 mm i.d. detection cell, respectively, with a white LED and a solid-state laser source. With an inexpensive 660 nm, 19 mW laser as the light source, the estimated detection limit for methylene blue (MB) was 30 nM, corresponding to 120 amol in the illuminated volume. The dimerization constant of MB and the quantum efficiency of the monomer was determined.
机译:我们在这里描述了一种基于在照明点的温度差和下游的温度差的流动流中的简单的光热检测方案。 我们使用单一双交叉25μm直径热电偶来测量温度变化。 黑暗中的基线标准偏差类似于0.001摄氏度,这取决于照明源增加到0.0016摄氏度。 我们通过1.5mm和0.1mm I.d,证明了几种色谱分离的染料的检测。 分别具有白色LED和固态激光源的检测单元。 由于廉价的660nm,19mw激光器作为光源,亚甲基蓝(Mb)的估计检测限为30nm,相当于照明体积的120 amol。 测定MB的二聚化常数和单体的量子效率。