Simplification in the Acquisition and Analysis of Fluorescence Decays Acquired with Polarized Emission for Time-Resolved Fluorescence Anisotropy Measurements

机译:荧光衰减的获取和分析中的简化,以时间分辨荧光各向异性测量获得

获取原文
获取原文并翻译 | 示例
           

摘要

This study introduces a global fluorescence decay analysis that substantially simplifies the acquisition and analysis of time-resolved fluorescence decays acquired with a vertically polarized excitation and vertically (I-VV(t)) and horizontally (I-VH(t)) polarized emission for time-resolved fluorescence anisotropy (TRFA) measurements. TRFA measurements were conducted whereby the I-VV(t) and I-VH(t) fluorescence decays of a series of oligoquinolines labeled at one end with an oligo(phenylenevinylene) dye (OPV-Q(n) with n = 4, 7, 17, 24, 33) were acquired according to the standard protocol that is currently accepted in the scientific literature which involves toggling the emission polarizer before fitting linear combinations of the I-VV(t) and I-VH(t) decays or acquiring the I-VV(t) and I-VH(t) decays with static polarizers before fitting them globally. The rotational time (phi) and initial anisotropy (r(0)) retrieved from these analyses were identical within experimental error regardless of whether the decays were acquired with toggling or static polarizers and fitted according to the standard protocol or globally. These experimental results were further supported by retrieving the parameters used to generate mono-, bi-, and tri-exponential TRFAs from the global analysis of simulated I-VV(t) and I-VH(t) fluorescence decays which were found to match perfectly the values that were inputted. Together, these experiments and simulations demonstrated that the parameters describing any type of TRFA can be extracted directly from the analysis of the I-VV(t) and I-VH(t) fluorescence decays acquired with a standard time-resolved fluorometer, a substantial simplification compared to the protocols currently in place to determine the TRFA.
机译:本研究介绍了全局荧光衰减分析,基本上简化了用垂直极化激发和垂直(I-VV(T))和水平(I-VH(T))极化发射的时间分辨荧光衰减的采集和分析时间分辨荧光各向异性(TRFA)测量。进行TRFA测量,由此在用寡核苷酸(苯基乙烯基)染料(OPV-Q(N)一端标记的一系列寡核苷酸的I-VV(T)和I-VH(T)荧光衰减(用N = 4,7,7 ,根据当前在科学文献中接受的标准方案获得了17,24,33),该方案涉及在拟合I-VV(T)和I-VH(T)衰减或获取的线性组合之前切换排放偏振器在全局拟合它们之前,I-VV(T)和I-VH(T)与静态偏振器衰减。从这些分析中检索的旋转时间(PHI)和初始各向异性(R(0))在实验误差内是相同的,而不管是否通过切换或静态偏振器获取衰减并根据标准协议或全球拟合。通过检索用于从模拟I-VV(t)和I-VH(T)荧光衰变的全局分析来产生用于产生单,双指数TRFA的参数来进一步支持这些实验结果。发现匹配的荧光衰变完美地输入的值。这些实验和模拟表明,可以直接从I-VV(T)和I-VH(T)荧光衰减的分析中直接提取描述任何类型的TRFA的参数,该荧光衰减是一个实质性的与目前用于确定TRFA的协议相比简化。

著录项

  • 来源
    《Analytical chemistry》 |2020年第1期|共6页
  • 作者单位

    Univ Waterloo Dept Chem Waterloo Inst Nanotechnol Polymer Res Inst 200 Univ Ave West Waterloo ON N2L 3G1 Canada;

    Univ Waterloo Dept Chem Waterloo Inst Nanotechnol Polymer Res Inst 200 Univ Ave West Waterloo ON N2L 3G1 Canada;

    Univ Waterloo Dept Chem Waterloo Inst Nanotechnol Polymer Res Inst 200 Univ Ave West Waterloo ON N2L 3G1 Canada;

    Univ Bordeaux CNRS Bordeaux Inst Natl Polytech CBMN UMR 5248 Inst Europeen Chim Biol 2 Rue Escarpit F-33600 Pessac France;

    Univ Bordeaux CNRS Bordeaux Inst Natl Polytech CBMN UMR 5248 Inst Europeen Chim Biol 2 Rue Escarpit F-33600 Pessac France;

    Univ Bordeaux CNRS Bordeaux Inst Natl Polytech CBMN UMR 5248 Inst Europeen Chim Biol 2 Rue Escarpit F-33600 Pessac France;

    Univ Bordeaux CNRS Bordeaux Inst Natl Polytech CBMN UMR 5248 Inst Europeen Chim Biol 2 Rue Escarpit F-33600 Pessac France;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 分析化学;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号