Several cerebral oximeters based on near-infrared spectroscopy (NIRS) are commercially available that determine tissue oxygen saturation (StO(2)). One problem is an inconsistency of StO(2) readings between different brands of instruments. Liquid blood phantoms mimicking optical properties of the neonatal head enable quantitative device comparisons. However, occasionally, the reduced scattering coefficient (mu(s)') of these phantoms decreases over time. Aim: To investigate whether this decrease in mu(s)' affects the validity of comparison of these devices. StO(2) was measured by several NIRS oximeters simultaneously on a phantom, which exhibited a particularly strong decrease in mu(s)'. We found that a decrease in mu(s)' by = 16% from baseline led to deviations in StO(2) of = 3%.
机译:基于近红外光谱(NIRS)的几种脑血管计是可商购的,可确定组织氧饱和度(STO(2))。 一个问题是STO(2)仪器之间的读数读数不一致。 液体血模模拟新生儿头的光学性质使定量装置比较能够进行定量。 然而,偶尔,这些幽灵的减小的散射系数(MU(S))随时间的推移降低。 目的:研究MU(S)'是否会影响这些设备比较的有效性。 STO(2)通过几个血管计同时测量了含有的幽明素,这表现出亩的特别强劲下降。 我们发现,通过&μS'的降低来自基线的μl(s)L导出为偏差在STO(2)中偏差。