Brief report: The sensitivity of motor responses for detecting catheter-nerve contact during ultrasound-guided femoral nerve blocks with stimulating catheters.

机译:简要报告:在超声引导的股神经阻滞刺激下,运动反应对于检测导管神经接触的敏感性。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: We determined the sensitivity of motor responses evoked by stimulating catheters in determining catheter-nerve contact using ultrasonography as reference. METHODS: Femoral nerves were contacted using stimulating catheters under ultrasonography scanning in 25 patients. The output current was increased from its minimum until quadriceps muscle contraction occurred. The sensitivity of the motor response in determining catheter-nerve contact was calculated using 0.5 mA as current threshold. RESULTS: The current required for catheter stimulation to evoke a motor response ranged between 0.18 and 2.0 mA. Muscle contraction in response to 0.5 mA occurred in 16 of 25 subjects. The sensitivity of motor response for nerve stimulation was 64% (95% confidence interval: 0.43, 0.82). CONCLUSIONS: The absence of muscle responses at a stimulating current
机译:背景:我们确定了通过超声检查作为参考,通过刺激导管引起的运动反应敏感性在确定导管-神经接触中的敏感性。方法:25例超声检查均采用刺激导管接触股神经。输出电流从最小值增加到四头肌肌肉收缩。使用0.5 mA作为电流阈值,计算确定导管-神经接触时电机响应的灵敏度。结果:导管刺激引起运动反应所需的电流范围为0.18至2.0 mA。 25位受试者中有16位发生了对0.5 mA的肌肉收缩反应。运动刺激对神经刺激的敏感性为64%(95%置信区间:0.43、0.82)。结论:在刺激电流为0.5 mA时没有肌肉反应并不一定表示没有导管神经接触。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号