首页> 外文期刊>Anesthesia and Analgesia: Journal of the International Anesthesia Research Society >A comparison of the classic and a modified laryngeal mask airway (OPLAC) in adult patients.
【24h】

A comparison of the classic and a modified laryngeal mask airway (OPLAC) in adult patients.

机译:成人患者经典喉罩和改良喉罩气道(OPLAC)的比较。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

BACKGROUND: A modified disposable laryngeal mask airway (LMA) (Oro-Pharyngo-Laryngeal Airway Cap, OPLAC) was developed in our department. In this study, we compared the performance of the LMA Classic with that of the OPLAC. METHODS: This was a randomized, single-blinded, crossover study involving 60 paralyzed, anesthetized adult patients. Both devices were inserted into each patient in different sequences after anesthesia had been induced. In 30 patients, the LMA was inserted first and in 30 patients, the OPLAC was inserted first. The success rate, insertion time, fiberoptic view, peak airway pressure, sealing pressure, incidence of gastric insufflations, trace of blood on the device, and incidence of postoperative sore throat were evaluated. RESULTS: The success rate of placement on the first attempt was high for both devices. The insertion time was significantly shorter and better engagement was noted on fiberoptic view with the OPLAC than with the LMA. The sealing pressure was significantly higher and the incidence of gastric insufflations was significantly lower with the OPLAC. The overall incidence of sore throat was 13.33%. CONCLUSIONS: Both devices have comparable airflow resistance and are easy to insert. The OPLAC requires less insertion time, has less variation on insertion time, fits better into the laryngopharynx, is less likely to cause gastric insufflations, and has a higher sealing pressure.
机译:背景:我们部门开发了一种改良的一次性喉罩气道(LMA)(Oro-Pharyngo-Langngeal气道帽,OPLAC)。在这项研究中,我们比较了LMA Classic和OPLAC的性能。方法:这是一项随机,单盲,交叉研究,涉及60例麻痹,麻醉的成年患者。麻醉后,将两种设备以不同的顺序插入每个患者。在30例患者中,首先插入LMA,在30例患者中,首先插入OPLAC。评估成功率,插入时间,光纤视图,峰值气道压力,密封压力,胃充气的发生率,装置上的血液痕迹以及术后喉咙痛的发生率。结果:两种设备首次尝试放置的成功率均很高。与LMA相比,OPLAC的插入时间明显更短,并且在光纤视图中观察到更好的接合。 OPLAC的密封压力显着较高,胃充气的发生率显着降低。喉咙痛的总发生率为13.33%。结论:这两种设备都具有相当的气流阻力,并且易于插入。 OPLAC需要较少的插入时间,插入时间的变化较小,更适合喉咽,不太可能引起胃气灌注,并具有较高的密封压力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号